Table 1.

Configurations of 8 SWEI excitations, Sequence #1 is similar to the energy used in current commercial SWEI implementations, Sequence #3 and #5 (bold) are a pair of high and low MI sequences that have the same excitation transmit energy (E), which was computed using Equation 2.

Seq Number MI Isppa(W/cm2) duration (µs) E (mJ)
1 1.6 420 364 4.0
2 1.6 420 546 6.0
3 1.6 420 727 8.0
4 1.6 420 1000 10.2
5 2.2 730 157 8.0
6 2.2 730 176 9.0
7 2.2 730 309 15.2
8 2.2 730 411 17.5