Skip to main content
. 2004 Jul;42(7):2961–2965. doi: 10.1128/JCM.42.7.2961-2965.2004

TABLE 4.

Estimates of sensitivity and specificity for HSV-2 assays grouped by city and by HIV status

Sensitivity or specificity and assay % Sensitivity or specificity in:
% Samples
Yaoundé Kisumu Ndola HIV HIV+
Sensitivity
    MAb EIA 97 98 100 95 100
    Gull Premier test 90 85 83 77 92
    Kalon Biological 92 94 89 89 94
    Biokit 89 89 77 83 89
    Radim 81 87 78 74 92
    Euroimmun 97 95 95 92 98
    IBL 97 98 68 90 96
    HerpeSelect 100 100 100 100 100
    DiaSorin 99 94 92 96 94
    Captia HSV-2 Type- Specific IgG 100 93 100 99 94
Specificity
    MAb EIA 98 98 94 98 88
    Gull Premier test 98 95 100 99 88
    Kalon Biological 98 95 100 99 88
    Biokit 90 92 97 95 81
    Radim 100 98 100 99 100
    Euroimmun 88 77 100 88 89
    IBL 83 50 100 76 100
    HerpeSelect 72 57 93 74 71
    DiaSorin 68 41 67 65 19
    Captia HSV-2 Type- Specific IgG 58 30 51 53 0