Table 2.

Performance parameters of the four RFA systems for 3 cm target ablation volume

Cool-Tip 1500X RF RF 3000 CelonPower LAB
Manufacturer Radionics AngioDynamics Boston Scientific Celon
x-axisa (cm) 4.0±0.5 3.3±0.3 3.7±0.6 4.0±0.5
y-axisb (cm) 3.7±0.4 3.0±0.6c 3.8±0.7c 3.5±0.4
x/y quotientd 1.09±0.14 1.14±0.23 0.98±0.18 1.17±0.16
z-axise (cm) 3.8±0.5 3.2±0.4 3.9±0.5 3.9±0.6
Volume (cm3) 28.5±6.5 17.1±4.9f 29.7±11.7f 28.8±7.0
Ablation timeg (min) 12.0±0.0 9.36±0.63 4.60±0.50 20.85±0.86
Temperatureg (°C) 72.5±14.9 60.8±10.9 61.6±22.0 54.2±15.0
P h < 0.001 0.135 0.007 0.001
Deviationi (%) 101 21 110 103

Overall P values for ANOVA are significant (P ≤ 0.043): i.e., x-, y-, and z-axis, and volume.

a

Diameter in the longitudinal axis along the electrode track.

b

Diameter in the perpendicular axis in the plane created by cutting along the electrode track.

c

P = 0.04 for AngioDynamics vs. Boston Scientific.

d

Degree of sphericity: the more spherical, the more the ratio approaches 1.

e

Diameter in the axis orthogonal to the x-y plane, at the midpoint of the x-axis on both sides, summated.

f

P = 0.03 for AngioDynamics vs. Boston Scientific.

g

Values for ablation time and temperature have descriptive character and were not used for ANOVA.

h

Double t test for paired samples: a value P < 0.05 implies a significant difference between the actual and the mathematically calculated volume.

i

Percent deviation of the actual volume from the mathematically calculated volume (14.14 cm3).

Data are presented as mean±standard deviation for the 3 cm RF-induced coagulation necrosis.