. 2014 Feb 7;20(3):251–258. doi: 10.5152/dir.2013.13157

Table 3.

Performance parameters of the four RFA systems for 5 cm target ablation volume

Cool-Tip 1500X RF RF 3000 CelonPower LAB
Manufacturer Radionics AngioDynamics Boston Scientific Celon
x-axisa (cm) 4.4±0.5b 4.5±0.6b 4.1±0.5b 5.4±0.5b
y-axisc (cm) 4.6±0.4d 4.2±1.0d 5.9±1.0d 5.3±0.7
Mean x/y quotiente 0.96±0.13 1.11±0.24 0.73±0.23 1.02±0.14
z-axisf (cm) 4.6±0.5g,h 3.9±0.5g,h 5.8±0.5h 6.1±1.1g
Volume (cm3) 48.3±9.9i 39.4±16.2i,j 71.8±14.5j 93.9±28.1i
Ablation timek (min) 12.0±0.0 19.59±1.13 9.15±2.93 40.21±1.78
Temperaturek (°C) 60.0±23.2 57.6±23.0 84.5±7.7 79.6±18.6
P l 0.002 0.003 0.25 0.024
Deviationm (%) 26 40 10 43

Overall P values for ANOVA are significant (P ≤ 0.004): i.e., x-, y-, z-axis, x/y quotient, and volume.

a

Diameter in the longitudinal axis along the electrode track.

b

P = 0.01 for Radionics vs. Celon; P = 0.001 for AngioDynamics vs. Celon; P < 0.001 for Boston Scientific vs. Celon.

c

Diameter in the perpendicular axis in the plane created by cutting along the electrode track.

d

P = 0.03 for Radionics vs. Boston Scientific; P = 0.003 for AngioDynamics vs. Boston Scientific.

e

Degree of sphericity: the more spherical, the more the ratio approaches 1.

f

Diameter in the axis orthogonal to the x-y plane, at the midpoint of the x-axis on both sides, summated.

g

P = 0.002 for Radionics vs. Celon; P < 0.001 for AngioDynamics vs. Celon.

h

P = 0.02 for Radionics vs. Boston Scientific; P < 0.001 for AngioDynamics vs. Boston Scientific.

i

P < 0.001 for Radionics vs. Celon; P < 0.001 for AngioDynamics vs. Celon.

j

P = 0.02 for AngioDynamics vs. Boston Scientific.

k

Values for ablation time and temperature measurement have descriptive character and were not used for ANOVA.

l

Double t test for paired samples: a value P < 0.05 implies a significant difference between the actual and the mathematically calculated volume.

m

Percentage deviation of the actual volume from the mathematically calculated volume (65.45 cm3).

Data are presented as mean±standard deviation for the 5 cm RF-induced coagulation necrosis.