Published in final edited form as: J Refract Surg. 2009 Nov 16;26(9):646–659. doi: 10.3928/1081597X-20091105-01

Table 1. Within-eye Point-repeatability and Region-repeatability of 10 Eyes That Underwent Artemis VHF Digital Ultrasound 1 Year After LASIK.

Epithelium Stroma Cornea Flap Residual Stromal Bed
Within-eye Point-repeatability – Corneal Vertex
Within-eye point-repeatability (μm) 0.58 [0.18, 0.99] 1.78 [0.40, 2.63] 1.68 [0.85, 2.32] 1.68 [0.85, 2.74] 2.27 [1.09, 4.28]
Coefficient of repeatability (μm) 1.14 3.50 3.29 3.29 4.46
Coefficient of variation (%) 0.97 0.43 0.35 1.39 0.64
Intraclass correlation coefficient 0.973 0.999 0.999 0.969 0.998
Within-eye Point-repeatability – Minimum Thickness
Minimum thickness repeatability (μm) 1.36 [0.67, 1.68] 2.76 [0.49, 6.08]
Within-eye Region-repeatability (μm)
Within 1-mm radius 1.01 [0.73, 1.35] 3.44 [2.32, 4.54] 3.35 [2.03, 4.33] 2.81 [2.01, 3.66] 3.97 [2.66, 5.30]
Within 2-mm radius 0.94 [0.70, 1.27] 4.92 [3.27, 6.46] 4.84 [3.00, 6.59] 2.79 [2.12, 3.50] 5.18 [3.50, 7.11]
Within 3-mm radius 1.04 [0.75, 1.32] 6.49 [4.25, 8.54] 6.02 [3.85, 7.86] 4.35 [2.94, 7.25] 7.58 [5.04, 10.87]

Note. The range for the study population is reported in brackets.