Skip to main content
. 2015 May 14;2015:496271. doi: 10.1155/2015/496271

Table 2.

Evaluation of the linearity of AOPPTIOC measurements.

Plasma sample
number
Dilution (%) Mean blank Mean test AOPPTIOC (μM chloramine T equivalents)
OD340 OD340 AOPPTIOC1a x dilution AOPPTIOC2b x dilution
1 5 0.068 0.032 4.37 87.4 2.09 41.8
10 0.123 0.131 17.17 171.7 4.14 41.4
20 0.209 0.602 78.38 391.9 8.85 44.3

2 5 0.075 0.040 5.35 107.1 2.31 46.3
10 0.130 0.161 21.07 210.7 4.59 45.9
20 0.224 0.641 83.46 417.3 9.14 45.7

3 5 0.123 0.058 7.76 155.2 2.79 55.7
10 0.220 0.202 26.48 264.8 5.15 51.5
20 0.404 0.800 104.13 520.6 10.20 51.0

OD340: optical density at 340 nm. aAOPPTIOC1: AOPPTIOC concentration calculated from original optical density values using the standard curve. bAOPPTIOC2: square root of AOPPTIOC1.