. Author manuscript; available in PMC: 2015 Jun 15.
Published in final edited form as: Phys Med Biol. 2005 May 5;50(10):2291–2311. doi: 10.1088/0031-9155/50/10/008

Table 3.

A comparison between optical properties (μa, μs, μeff) determined by broad-beam and parallel-catheter methods for three optical phantoms with Liposyn concentrations: (A) 0.23%, (B) 0.53%, (C) 1.14% and ink concentrations (0.002%, 0.012% and 0.023%) for three different separations (h) between the catheters.

Parallel catheters Broad beam Difference (%)



Separation h μa
μs
μeff μa
μs
μeff μa
μs
μeff
(A)
h = 3 mm 2.5 0.10 2.04 0.77 0.10 1.73 0.72 2.0 14.0 6.9
0.47 2.11 1.72 0.48 1.73 1.58 5.1 17.9 7.5
0.93 2.49 2.64 1.00 1.73 2.28 8.6 39.1 15.8
h = 5 mm 4.5 0.11 2.04 0.82 0.10 1.73 0.72 8.0 17.9 13.9
0.53 2.21 1.87 0.48 1.73 1.58 9.4 27.7 18.4
1.10 2.29 2.75 1.00 1.73 2.28 9.9 32.4 20.6
h = 7 mm 6 0.11 1.93 0.79 0.10 1.73 0.72 9.0 7.8 9.7
0.55 1.91 1.78 0.48 1.73 1.58 12.7 6.7 11.3
1.18 1.49 2.29 1.00 1.73 2.28 15.2 16.8 0.4
(B)
h = 3 mm 2.5 0.10 4.35 1.15 0.10 4.19 1.11 1.0 3.8 3.6
0.43 4.74 2.47 0.49 4.19 2.49 12.9 13.1 0.8
0.89 4.30 3.38 0.99 4.19 3.53 10.3 2.6 4.2
h = 5 mm 4.5 0.09 3.92 1.05 0.10 4.19 1.11 7.0 6.4 5.4
0.56 3.63 2.46 0.49 4.19 2.49 13.5 13.4 1.2
1.14 3.53 3.47 0.99 4.19 3.53 15.1 15.8 1.7
h = 7 mm 6 0.08 3.35 0.92 0.10 4.19 1.11 16.0 20.0 17.1
0.59 2.80 2.23 0.49 4.19 2.49 20.6 33.2 10.4
1.35 2.57 3.23 0.99 4.19 3.53 36.5 38.7 8.5
(C)
h = 3 mm 2.5 0.09 6.97 1.35 0.1 9.14 1.64 13.0 23.7 17.7
0.46 5.05 2.63 0.5 9.14 3.7 8.8 44.7 28.9
0.81 5.03 3.49 0.99 9.14 5.2 18.4 45.0 32.9
h = 5 mm 4.5 0.1 8.76 1.62 0.1 9.14 1.64 0 4.2 1.2
0.51 7.40 3.38 0.5 9.14 3.7 2.0 19.0 8.6
0.92 8.02 4.70 0.99 9.14 5.2 7.3 12.3 9.6
h = 7 mm 6 0.10 5.92 1.32 0.1 9.14 1.64 2.0 35.2 19.5
0.78 3.97 3.04 0.5 9.14 3.7 55.6 56.6 17.8
1.85 3.27 4.26 0.99 9.14 5.2 86.9 64.2 18.1