TABLE III.

Passing Rate Difference Between Human Fit Factor (FF) and Static Headform Manikin Fit Factor (MFF) by Test Subject

Subject ID Total Tests (n) Test Type Fail (n) Pass (n) Pass (%)A Pass Rate Difference (%)B
1 24 Human FF 15 9 37.5 −12.5
24 StAH MFF 12 12 50.0
2 24 Human FF 18 6 25.0 −37.5
24 StAH MFF 9 15 62.5
3 16 Human FF 5 11 68.8 18.8
16 StAH MFF 8 8 50.0
4 24 Human FF 8 16 66.7 4.2
24 StAH MFF 9 15 62.5
5 21 Human FF 9 12 57.1 0.0
21 StAH MFF 9 12 57.1
6 16 Human FF 5 11 68.8 25.0
16 StAH MFF 9 7 43.8
7 24 Human FF 7 17 70.8 16.7
24 StAH MFF 11 13 54.2
8 24 Human FF 9 15 62.5 8.3
24 StAH MFF 11 13 54.2
9 24 Human FF 9 15 62.5 0.0
24 StAH MFF 9 15 62.5
10 24 Human FF 6 18 75.0 12.5
24 StAH MFF 9 15 62.5
A

Passing criterion is FF or MFF result ≥ 100. Pass (%) = Pass (n) / Total tests (n) x 100%.

B

Pass rate difference (%) = Human FF Pass (%) – StAH MFF Pass (%).