Skip to main content
. Author manuscript; available in PMC: 2016 Jul 7.
Published in final edited form as: Lab Chip. 2015 May 26;15(13):2790–2798. doi: 10.1039/c5lc00305a

Table 1.

Throughput and sensitivity of the sorter

Genotype us
(μm/s)
u3
(μm/s)
Total
number
of
worms
% in
initial
mix
Sorting
time (h)
% in
holding
chamber
% in
collection
chamber
Throughput
(#/h)
unc-63 40 165 496 50 1.6 90.9 0.7 317
dys-1 160 204 991 50 1.1 90.0 11.7 901
lev-10 121 127 1036 50 0.4 96.0 10.7 2486
Average 92.3 7.7 1235

The % are for the slow mutants, which lacked fluorescent protein expression.

us: Swimming velocityu3: Fluid velocity in the separation conduit