Skip to main content
. 2015 Jun 18;53(7):2072–2078. doi: 10.1128/JCM.00157-15

TABLE 1.

GM-EIA OD characteristics for testing of serum samples and threshold controls by manual and DS2 platform processing

Parameter Value
Overalla (n = 121)
Negative samples (n = 101)
Spiked positive samples (n = 20)
Threshold controls (n = 10)
Manual DS2 platform 1 DS2 platform 2 Manual DS2 platform 1 DS2 platform 2 Manual DS2 platform 1 DS2 platform 2 Manual DS2 platform 1 DS2 platform 2
Range 0.023–1.44 0.055–1.876 0.04–2.693 0.023–0.149 0.055–0.299 0.04–0.329 0.227–1.440 0.486–1.876 0.313–2.693 0.378–0.687 0.742–1.106 0.603–1.135
25th percentile 0.04 0.076 0.079 0.037 0.073 0.075 0.442 0.807 0.688 0.42 0.88 0.68
Median OD 0.048 0.106 0.109 0.045 0.093 0.102 0.627 1.058 1.141 0.48 0.76 0.79
75th percentile 0.078 0.151 0.159 0.057 0.127 0.127 0.912 1.341 1.580 0.53 1.0 1.1
a

Does not include threshold controls.