Table 1. Detailed instrument properties.

OCT CRS
Illumination power   3.0 mW at sample plane   30.0 mW at sample plane
Illumination wavelength   λ0 = 830 nm, Δλ = 45 nm   λ0 = 785 nm, Δλ = 0.5 nm
SNR   94 dB   36 dB
Axial resolution (in air)   8 µm   19 µm
Lateral resolution (in air)   40 µm   3 µm
Spectral resolution   0.07 nm   3.3 cm−1
Spectral range   810 – 860 nm   150 – 1950 cm−1
OCT imaging depth   1.6 mm (−3 dB falloff)   -
OCT field of view   900 µm   -

A description of the procedures used to measure these properties can be found in section 2.1.