Skip to main content
. Author manuscript; available in PMC: 2015 Jun 19.
Published in final edited form as: Phys Med Biol. 2012 Nov 6;57(23):7783–7797. doi: 10.1088/0031-9155/57/23/7783

Table 1.

Average relative uncertainty ( σ/D¯), γ-test passing rates (Pt ) with different criteria and computation time (T) for all different test cases.

Phantom Source energy (MeV) σ/D¯ (%) Pγ (1 mm/1%) (%) Pγ (2 mm/2%) (%) T (s)
Water, EM 100 0.8 99.2 99.6 6.66
200 0.8 99.9 99.9 19.12
Water 100 0.9 99.3 99.7 6.76
200 1.1 97.3 99.9 21.14
Bone 100 0.9 98.6 98.7 6.68
200 1.1 98.1 99.9 20.98
Tissue 100 0.9 99.0 99.4 7.08
200 1.1 97.6 99.9 22.29
Inhomogeneous phantom 100 0.9 99.9 99.9 9.52
Patient 100 1.0 95.1 99.9 10.08