Table 1.

Optimized 96-Well High-Throughput Screening Parameters and Assay Statistics.a

SAMHD1 0.5 μM
PPase 5 μM
dGTP (activator/substrate) 0.1 mM (~1/10 KmdGTP)b
Total reaction volume and time 40 μL and 20 min
Z′ factor 0.87
Mean signal (± SD)
 Positive controlc 2.28 ± 0.08
 Negative controld 0.26 ± 0.01
Coefficient of variation of signal, %
 Positive control 3.4
 Negative control 3.3
Signal/backgrounde 9
a

The volumes would be reduced by 50% for a 384-well plate format.

b

See reference 4 for the steady-state kinetic parameters of SAMHD1.

c

The mean measured absorbance for 96 reactions carried out using the indicated conditions.

d

The mean measured absorbance for 96 reactions carried out in the absence of SAMHD1.

e

Calculated from the ratio of the mean signals for the positive and negative controls.