Skip to main content
. 2015 Jun 8;2015:109353. doi: 10.1155/2015/109353

Table 4.

Yield and viability of SVF obtained by the automated process using filtration technology as compared to the manual process of cell isolation using centrifugation.

Digestion Filtration
Manual Automated Manual Automated
SVF yield (×105/gram) 0.90 ± 0.5 0.82 ± 0.6 1.15 ± 0.3 1.17 ± 0.5
Percentage yield 100% 94 ± 28% 100% 98 ± 21%
Percentage viability 96 ± 3.0% 96 ± 2.1% 97.3 ± 1.5% 97.5 ± 2.8%

Data is represented as mean ± SD for five individual samples tested. SVF Yield is represented as cell number ×105/gram of adipose tissue. Percentage yield indicates percentage recovery by automated processing relative to the yield from manual processing taken as the baseline comparator. Difference in cell yield between manual centrifugation method and automated filtration is not significant (p = 0.8, paired t-test).