. 2015 May 1;17(4):976–987. doi: 10.1208/s12248-015-9762-4

Fig. 4.

Fig. 4

ac Assay performance and stability of pre-coated and StabilCoat-treated Streptavidin plates. Standard curves (a) and QC performance (b) from analytical runs using fresh plates and pre-coated Streptavidin plates. The fresh plates were coated with the capture antibody on the day of the assay. The pre-coated plates were coated with capture antibody, treated with 0.5X or 1X StabilCoat and kept at 4°C or 37°C for 48 h (no differences between the 4°C and 37°C groups and the data were grouped together, n = 4). Assay QC performance using 0.5X StabilCoat-treated plates at a CRO were also presented (n = 18). (c) Accelerated stability study of pre-coated Streptavidin plates. QC performance in analytical runs using fresh or pre-coated plates (treated with 0.5X StabilCoat) stored at 37°C for 2 weeks. Results are shown as mean and SD (n = 4)