Skip to main content
. Author manuscript; available in PMC: 2015 Jun 23.
Published in final edited form as: Biotechniques. 2013 May;54(5):257–264. doi: 10.2144/000114028

Table 2.

Comparison of the performance of multiplexed antibody assay without and with MiNC

Name LOD (ng/ml) Dynamic range (ng/ml) Intra-CV(%)* Inter-CV(%)*
MiNC MiNC MiNC MiNC
+ + + p-value + p-value
CYRAB 0.4 0.4 0.4 – 2000 0.4 – 2000 9 3 <0.001 15 10 0.278
c-Jun 57.2 22.2 57.2 – 40000 22.2 – 40000 9 3 <0.001 6 10 0.235
PRDX4-3 5.6 3.4 5.6 – 20000 3.4 – 20000 12 4 <0.001 10 10 0.944
AV 10 4 10 10
*

The intra-CV and inter-CV were calculated as the average CV across eight antibody concentrations.