Spatial accuracy of the CT and the MRI scanners.

Imager   Zone 1 |ΔE|
Zone 2 |ΔE|
Zone 3 |ΔE|
Zone 4 |ΔE|
Exact (mm) 15.56 31.11 62.23 93.34
O-arm Point 1 15.60 0.04 31.11 0.002 62.13 0.09 93.33 0.01
Point 2 15.59 0.03 31.19 0.08 62.22 0.01 93.33 0.01
Point 3 15.42 0.14 31.28 0.17 62.13 0.10 93.25 0.09
Point 4 15.45 0.11 31.19 0.08 62.05 0.18 92.99 0.35
Average displacement (mm) 15.52 0.08 31.19 0.08 62.13 0.10 93.22 0.12
±SD (mm) 0.05 0.03 0.04 0.03 0.04 0.04 0.08 0.08
GE Signa Point 1 15.67 0.11 31.17 0.06 62.42 0.19 93.16 0.18
Point 2 15.58 0.02 31.17 0.06 62.34 0.11 93.59 0.26
Point 3 15.67 0.11 31.08 0.03 62.59 0.37 94.03 0.59
Point 4 15.41 0.15 30.82 0.29 62.16 0.07 93.16 0.18
Average displacement (mm) 15.48 0.10 31.06 0.11 62.38 0.19 93.48 0.30
±SD (mm) 0.06 0.03 0.08 0.06 0.09 0.06 0.21 0.10

Note: Exact=known coordinate values at each point; |ΔE|=absolute values of displacement from the exact points on the phantom; Point 1, Point 2, Point 3, and Point4=each measured point on the phantom. The points are in the phantom as follows: Point 1 is at the corner closest to the Roman numeral indicating each zone followed by the other points, in turn, at the corner of each zone clockwise from Point 1; SD=standard deviation.