. 2015 May 13;42(8):1202–1222. doi: 10.1007/s00259-015-3059-9

Table 5.

Reconstruction parameters that provided most accurate quantification over the assessments performed in this investigation, and the expected accuracy and standard deviations associated with measures of warm background and hot spheres for each. All measures are based on the results from day-0 imaging, where phantom background and hot sphere concentration were about 300 kBq/ml and 2,500 kBq/ml, respectively

Vendor Model Recommended reconstruction for quantitative purposes Error in warm background concentration measures (%) Error in 37-mm hot sphere concentration measures (%)
Average ± SD Range Average ± SD Range
GE Healthcare Discovery 600 Discovery ST (E) Discovery RX 3D OSEM with all-pass filter: e.g. 2i24s −9 ± 10 +4 – 29 −34 ± 9 −14 – 49
GE Healthcare Discovery 690 Discovery 710 3D OSEM with all-pass filter: e.g. 2i24s + RR + ToF 1 ± 4 +6 – 7 −14 ± 9 −5 – 28
Philips Gemini TF 3D OSEM (BLOB OS TF) with no filter: 4i8s + ToF −5 ± 2 −4 – 6 −22 ± 3 −20 – 24
Siemens Biograph (various) 3D OSEM with all-pass filter: e.g. 2i21s + RR (acquired in ‘PROMPTS + RANDOMS’ mode) 2 ± 9 +9 – 22 −27 ± 5 −20 – 40
Siemens Biograph mCT 3D OSEM with all-pass filter: 2i21s + RR + ToF −2 ± 6 +4 – 9 −16 ± 4 −13 – 22