Table 5.
Vendor | Model | Recommended reconstruction for quantitative purposes | Error in warm background concentration measures (%) | Error in 37-mm hot sphere concentration measures (%) | ||
---|---|---|---|---|---|---|
Average ± SD | Range | Average ± SD | Range | |||
GE Healthcare | Discovery 600 Discovery ST (E) Discovery RX | 3D OSEM with all-pass filter: e.g. 2i24s | −9 ± 10 | +4 – 29 | −34 ± 9 | −14 – 49 |
GE Healthcare | Discovery 690 Discovery 710 | 3D OSEM with all-pass filter: e.g. 2i24s + RR + ToF | 1 ± 4 | +6 – 7 | −14 ± 9 | −5 – 28 |
Philips | Gemini TF | 3D OSEM (BLOB OS TF) with no filter: 4i8s + ToF | −5 ± 2 | −4 – 6 | −22 ± 3 | −20 – 24 |
Siemens | Biograph (various) | 3D OSEM with all-pass filter: e.g. 2i21s + RR (acquired in ‘PROMPTS + RANDOMS’ mode) | 2 ± 9 | +9 – 22 | −27 ± 5 | −20 – 40 |
Siemens | Biograph mCT | 3D OSEM with all-pass filter: 2i21s + RR + ToF | −2 ± 6 | +4 – 9 | −16 ± 4 | −13 – 22 |