Technical specifications of the CO2 measuring devices used in this study.

NDIR Sensor OP-Laser PAS Analyzer
Operating principle Non-Dispersive Infra-Red Tunable Diode Laser Absorption Spectroscopy Photoacoustic spectroscope
Target gas CO2 only CO2 only CO2 and others
Tuned wavelength for detection of CO2 (µm) 4.3 1.6 14.1
Detection limit 0.2 ppmv 2000 ppm∙m 1.5 ppmv
Accuracy ±30 ppmv ± 2% 500 ppm∙m ±1%
Response time (T1/e) 20 s (diffusion time) 1 s (scan rate) 25 s (for 1 gas, including flushing period and 5 s sampling integration time)
Operating temperature (°C) 0 to +50 °C −45 °C to +80 °C −20 °C to +70 °C
Sensitivity to pressure 1 0.1% Not determined 0.5%
Type of calibration Two-points calibration in the laboratory Continuous self-calibration mechanism with a reference gas crystal cell Two-points calibration at factory
Laboratory check of calibration with certified gas Yes Yes Yes
Cross-compensation for water vapor Not needed Not needed Yes
Cost per unit (€) 2 300 35,000 28,000

1 Specified by factory; 2 Approximate, excluding costs with external datalogger systems, when applicable.