Table 1.

Comparison of absorbance results from LOCD ELISA detection platform with absorbance values from spectrophotometer.

Mean OD of LOCD Reader at 450 nm Mean OD of Spectrometer at 445 nm–455 nm Error Tolerance (%)
3.447 3.586 −0.14 −3.9
2.634 2.956 −0.323 −10.9
2.367 2.636 −0.269 −10.2
2.11 2.236 −0.126 −5.6
1.738 1.705 0.033 1.9