Skip to main content
. Author manuscript; available in PMC: 2015 Jun 26.
Published in final edited form as: Rep Prog Phys. 2010 Jun 2;73(7):076701. doi: 10.1088/0034-4885/73/7/076701

Figure 4.

Figure 4

Common geometry used to model tissue. In the semi-infinite geometry, w, h, and d all go to infinity, while in the infinite slab geometry, w and h are infinite but d is finite. Both geometries have azimuthal symmetry about the z-axis, meaning the photon fluence rate only depends on the radial and axial cylindrical coordinates ρ and z. The unit vector points from inside the tissue to outside. On the left, a single source-detector pair (with separation ρ) in the reflection geometry is shown. Note that for the slab geometry, detectors can also be used for transmission measurements by being placed on the z = d plane. On the right is a cross-section showing that the radiance moving into the turbid medium at the boundary is due to the Fresnel reflection of the radiance incident on the boundary.