Skip to main content
. 2014 Oct 6;119:554–574. doi: 10.6028/jres.119.022

Table 7.

Adopted correction factors for measurements made with the NIST FACs for the Xoft Axxent source at 50 kV.

Factor For: Lamperti Ritz
1 kion ion recombination ≈1.0000 ≈1.0000
2 khumidity humidity of air 0.998 0.998
3 katt Attenuation 1.0087 1.0283
4 kel electron loss 1.0008 1.0000
5 ksc photon scatter 0.9987 0.9970
6 kfl fluorescence reabsorption 0.9979 0.9969
7 kbr/(1−g) effects of bremsstrahlung 1.0 1.0
8 kii initial ion 1.0 1.0
9 kdia diaphragm scatter 1.0 1.0
П k1−9 1.0041 1.0200
Kvac/Kair conversion to air-kerma strength 1.12 1.12