Table 3.

HPGe full-energy-peak efficiency measurements and applied correction factors for the T-detector

Source type Gamma-ray peak energy (keV) Efficiency Efficiency uncertainty (%) Self-attenuation correction factor Solution fill height correction factor
2 mL cylinders 276.3989 1.100E-04 0.616 0.915 0.985
302.8508 1.026E-04 0.875 0.918 0.985
356.0129 8.928E-05 0.309 0.923 0.985
383.8485 8.363E-05 0.553 0.926 0.985
4 mL cylinders 276.3989 1.092E-04 0.565 0.915 0.985
302.8508 1.008E-04 0.652 0.918 0.985
356.0129 8.827E-05 0.270 0.923 0.985
383.8485 8.265E-05 0.438 0.926 0.985
6 mL cylinders 276.3989 1.094E-04 0.452 0.915 0.985
302.8508 1.016E-04 0.449 0.918 0.985
356.0129 8.902E-05 0.242 0.923 0.985
383.8485 8.325E-05 0.362 0.926 0.985
23 mL cylinders 276.3989 1.124E-04 1.551 0.915 0.985
302.8508 1.035E-04 1.262 0.918 0.985
356.0129 9.074E-05 1.180 0.923 0.985
383.8485 8.498E-05 1.168 0.926 0.985