Table 3. Diagnostic performance of the newly developed ICT and the Kalazar Detect Rapid Test in the detection of VL.
Diagnostic parameter | Newly developed ICT | Kalazar Detect |
---|---|---|
True positive (tp) | 204 | 202 |
True negative (tn) | 231 | 229 |
False positive (fp) | 3 | 5 |
False negative (fn) | 9 | 11 |
Sensitivity a | 95.8% (95% CI: 92.1–98.1%) | 94.8% (95% CI: 91.0–97.4%) |
Specificity b | 98.7% (95% CI: 96.3–99.7%) | 97.9% (95% CI: 95.1–99.3%) |
Diagnostic efficiency c | 97.3% (95% CI: 95.4–98.6%) | 96.4% (95% CI: 94.3–97.9%) |
a Sensitivity = tp × 100/(tp + fn)
b Specificity = tn × 100/(tn + fp)
c Diagnostic efficiency = (tn + tp) × 100/(tp + fp + tn + fn)