TABLE 3.
Lateral Patient Width, cm | Noise Index | Rotation Time, s | Minimum Tube Current, mA | Maximum Tube Current, mA |
---|---|---|---|---|
22.1–30 | 9 | 0.5 | 150 | 280 |
30.1–40 | 11.5 | 0.5 | 220 | 500 |
40.1–45 | 14.5 | 0.5 | 400 | 720 |
45.1–50+ | 17 | 0.7 | 450 | 770 |
These values are for 5-mm slice thicknesses. A gantry rotation of 0.5 seconds is used until patient size exceeds 45 cm, when a gantry rotation of 0.7 seconds is required to achieve sufficient tube output.