Published in final edited form as: J Comput Assist Tomogr. 2015 May-Jun;39(3):437–442. doi: 10.1097/RCT.0000000000000221

TABLE 3.

Noise Index Technique Chart Used in Our Adult Abdominal CT Practice for an AEC System That Uses the Constant Noise Paradigm (VCT; GE Healthcare)

Lateral Patient Width, cm Noise Index Rotation Time, s Minimum Tube Current, mA Maximum Tube Current, mA
22.1–30 9 0.5 150 280
30.1–40 11.5 0.5 220 500
40.1–45 14.5 0.5 400 720
45.1–50+ 17 0.7 450 770

These values are for 5-mm slice thicknesses. A gantry rotation of 0.5 seconds is used until patient size exceeds 45 cm, when a gantry rotation of 0.7 seconds is required to achieve sufficient tube output.