Skip to main content
. 2015 Jul 10;23(2):178–184. doi: 10.1089/ars.2015.6262

Table 4.

Ultra-Performance Liquid Chromatography–Tandem Mass Spectrometry Acquisition Parameters for the Determination of Isoprostanes, Isofurans, Neuroprostanes, and Neurofurans in Urine of Preterm Infants in the First 4 Weeks After Birth

Analyte m/z Parent ion m/z Daughter ion Cone (V) Collision (eV) Time window (min)
IsoPs 353.2 115 40 20 3.9–6.0
IsoFs 369.2 115 45 20 3.9–6.0
NeuroPs 377 101 35 20 3.8–6.0
NeuroFs 393 193 35 20 3.8–6.0
PGF-D4 357.5 197.3 45 30 4.6–5.3