Figure 7.
(a) CAD model of NiTi specimen for static and fatigue characterization. The significant dimension of gauge length (L), wires width (W) and thickness (Th) are reported. (b) Symmetrical FE model of a tensile test up to 5%: the total maximum strain, uniform and distributed along each wire, is equal to 4.459%, which means that almost the whole load applied to the specimen is transferred on its gauge length.