Skip to main content
. 2015 May 20;6(2):299–317. doi: 10.3390/jfb6020299

Table 1.

Cyclic axial tests on peripheral stents: testing conditions in terms of mean and alternate displacement values (um, ua) expressed in mm and final results after 107 cycles.

Tests 8 Rings 4 Rings
um ua Fracture um ua Fracture
Test 1 11.05 ±2.55 Yes 5.525 ±1.275 Yes
Test 2 8.4 ±2.2 Yes 4.2 ±1.1 Yes
Test 3 8.55 ±2.55 Yes 4.275 ±1.275 Yes
Test 4 11.5 ±2 No 5.75 ±1 Yes
Test 5 −2.39 ±2.15 Yes 6 ±1.28 Yes
Test 6 4 ±0.25 No