Skip to main content
. 2015 Jun 20;7(6):e280. doi: 10.7759/cureus.280

Table 3. Comparison of US probe tracking techniques available in radiotherapy.

  Optical Electromagnetic Mechanical X-ray
Typical spatial resolution 0.2-0.4 mm 0.5-0.8 mm 0.1-5.0 mm 0.2 mm
Typical acquisition rate 20-60 Hz 20-200+ Hz 300-1000 Hz <0.2 Hz. Limited by safety considerations.
Line of sight issues? Yes Yes for some materials No Yes for some materials
Metal interference? None High None None
Ionizing radiation? No No No Yes