Skip to main content
. 2014 Oct 14;9(1):149–152. doi: 10.1177/1932296814554415

Table 1.

Glucose Meter Long-Term Biases.

Reference Meter Bias Percentage Bias
50 46.7 −3.4 −6.7
100 94.2 −5.8 −5.8
150 141.8 −8.3 −5.5
200 189.3 −10.7 −5.4
250 236.9 −13.2 −5.3
300 284.4 −15.6 −5.2
350 332.0 −18.1 −5.2
400 379.5 −20.5 −5.1