Published in final edited form as: Phys Med Biol. 2015 Jun 4;60(12):4753–4775. doi: 10.1088/0031-9155/60/12/4753

Table 1.

The measurement conditions for the profiles used to train the dose model are listed below including the field size, source-to-surface distance (SSD), depth, source-to-measurement-plane distance (SPD), and scan direction.

Water-Phantom

Field Size (cm2) SSD (cm) Depth (cm) Profile Source

5×5 100 1.5 In-Plane (Kaderka et al., 2012)
Cross-Plane
10 In-Plane
Cross-Plane This Worka
10×10* 100* 1.5* In-Plane* (Kaderka et al., 2012)
Cross-Plane

In-Air

Field Size (cm2) SPD (cm) Profile Source

5×5 101.5 In-Plane This Work
Cross-Plane
110 In-Plane
Cross-Plane
10×10* 101.5* In-Plane*
Cross-Plane
40×40 85 In-Plane
Cross-Plane
100 In-Plane
Cross-Plane
115 In-Plane
Cross-Plane
130 In-Plane
Cross-Plane

The measurement conditions indicated by * represent the reference conditions selected for this study.

a

Chauvenet’s Criterion was used to exclude the results of the water-phantom profile measured in this work due to the significantly lower dose resolution than those from Kaderka et al.