Table 3.

Parameters to model the absorbed dose from primary photon fluence in water.

Symbol Description In-Plane Cross-Plane Source
αQ1 Incident electron charge empirical adjustment parameter for Gaussian 1 1.24 1.20 This Worka
αQ2 Incident electron charge empirical adjustment parameter for Gaussian 2 0.62 1.00
ασ1 Width parameter adjustment factor for Gaussian 1 0.79 0.52
ασ2 Width parameter adjustment factor for Gaussian 2 0.51 0.74
α2 Adjustment parameter for the mean of Gaussian 2 0.82 1.51
αw Adjustment parameter to the μt of water 0.77
d0 Upstream distance from the effective photon source to the virtual photon source (mm) 6.03
μw Linear attenuation coefficient of water for 2 MeV photons (1/cm) 4.942 × 10−2 (Hubbell and Seltzer, 2011)
SSD Source-to-surface-distance (cm) 100 100 (Kaderka, 2012)
SDD Source-to-diaphragm-distance (cm) 47 37.7 (Elekta, 2007)
a

From iterative fitting procedure