Published in final edited form as: Phys Med Biol. 2015 Jun 4;60(12):4753–4775. doi: 10.1088/0031-9155/60/12/4753

Table 4.

Parameters to model the absorbed dose from un-attenuated leakage photon fluence.

Symbol Description In-Plane Cross-Plane Source
αPC,σ Adjustment parameter for source width for the leakage fluence attenuated by the primary collimator 0.87 4.13 This Worka
q Divergence factor for the leakage photon fluence 1.58 1.11
αSC Adjustment factor to the μt of the secondary collimator 1.46 1.85
αPC Adjustment factor to the μt of the primary collimator 0.32 0.28
SIDPC Distance from the focus of the primary collimator cone to isocenter (cm) 112
αSID Adjustment factor for the SIDPC in each plane 1.09 0.87
μSC Linear attenuation coefficient of the secondary collimator (1/cm) 0.77 0.58 (NCRP, 2005)
tSC Secondary collimator thickness (cm) 7.8 11.2 Personal Communication, Elekta, Stockholm, Sweden
μwSC Effective attenuation of the primary collimator 0.64 (NCRP, 2005), Personal Communication, Elekta, Stockholm, Sweden
PC Projected location of the primary collimator in the isocentric plane (cm) 25 Personal Communication, Elekta, Stockholm, Sweden
a

From iterative fitting procedure