. 2015 Jul 10;42(8):4579–4590. doi: 10.1118/1.4926428

TABLE I.

Summary of geometry-based evaluations of accuracy of the various methods in reconstructing the total infarct for the 36 human LGE-CMR images. An asterisk before a number indicates a statistically significant difference between the corresponding method and the LogOdds method.

Method DSC (%) δRV (%) RMSE (mm) δRSA (%) Bh Euler δχ
LogOdds 82.10 ± 6.58 7.23 ± 6.2 1.08 ± 0.15 7.29 ± 5.58 0.89 ± 0.02 6.25 ± 4.94
SDM 65.43 ± 12.86 48.77 ± 14.3 1.34 ± 0.21 28.15 ± 11.3 0.90 ± 0.01 11.6 ± 12.37
MSDM 75.99 ± 8.57 25.67 ± 10.15 1.20 ± 0.28 12.22 ± 8.94 0.90 ± 0.02 9.12 ± 11.83
VI 60.35 ± 17.36 96.32 ± 60.66 4.31 ± 3.36 18.02 ± 17.5 0.89 ± 0.02 5.36 ± 5.1
NN 74.13 ± 6.90 8.1 ± 14.56 1.54 ± 0.19 9.94 ± 4.7 0.68 ± 0.02 12.51 ± 8.10