Table 1.
Results of the iterative source path correction
Applicator | R34 | R30 | R26 |
---|---|---|---|
r TPS (mm) | 17 | 15 | 13 |
d(X obs – X TPS) ± 1σ (mm) | 3.2 ± 0.9 | 4.5 ± 1.3 | 3.6 ± 1.3 |
d max (X obs – X TPS) (mm) | 4.6 | 6.1 | 5.7 |
Fixed dwell position | 9 | 7 | 6 |
r corr (mm) | 17.50 | 15.85 | 13.95 |
d(X obs – X corr) ± 1σ (mm) | 0.6 ± 0.3 | 0.7 ± 0.4 | 0.8 ± 0.5 |
d max (X obs – X corr) (mm) | 1.3 | 1.3 | 2.0 |
d r (X obs – X corr) ± 1σ (mm) | 0.1 ± 0.7 | 0.1 ± 0.7 | –0.1 ± 0.8 |