Table 7. Effect size of the change in LCI after reanalysis with different algorithms for signal processing and detection limits.

Settings Changed LCI % change of standard LCI # children with 5–10% change # children with >10% change
Baseline 8.55 ± 2.64 [6.29–14.94]
O2 drift correction off 8.29 ± 2.18 [6.38–14.95] -1.75 ± 7.95 5 (17%) 5 (17%)
Dynamic CO2 correction off 9.07 ± 2.60 [6.65–15.99] 6.67 ± 5.23 15 (50%) 5 (17%)
Integrated reinspired N2 off 8.51 ± 2.57 [6.25–14.76] -0.41 ± 1.65 0 0
O2 response-time correction off 9.24 ± 2.71 [6.63–18.05] 8.65 ± 6.65 14 (47%) 11 (37%)
Volume Sensitivity higher 8.57 ± 2.63 [6.25–14.94] 0.23 ± 0.64 0 0

The table shows mean ± SD [range] LCI after reanalysis of the tests with changed default algorithms for signal correction, mean ± SD change in percentage of standard LCI, number of children with a change between 5–10% and children with >10% change of standard LCI.