Published in final edited form as: Clin Nephrol. 2009 Jul;72(1):5–14. doi: 10.5414/cnp72005

Table 3.

Cut-off thresholds (pg/ml) that achieve at least 80% specificity for predicting bone turnover abnormalities with corresponding sensitivities (Sens.), positive predictive values (PPV), and negative predictive values (NPV).

Assay Low High
Cut-off Sens. PPV NPV Cut-off Sens. PPV NPV
Diasorin 165 0.68 0.69 0.79 404 0.52 0.61 0.76
Elecsys 291 0.68 0.69 0.79 759 0.60 0.65 0.79
Centaur 355 0.74 0.72 0.82 1003 0.49 0.59 0.75
Immulite 550 0.69 0.72 0.80 979 0.51 0.61 0.77
Scantibodies 223 0.58 0.66 0.74 713 0.52 0.61 0.76