Skip to main content
. 2015 Jul 14;10(7):e0132929. doi: 10.1371/journal.pone.0132929

Table 1. Vendor-specific parameters of the pCASL product sequences.

GE Philips
Readout sequence 3D FSE stack-of-spirals 2D gradient-echo single-shot EPI
Acquisition matrix 8 arms with 512 sampling points 80 * 80
Parallel imaging No SENSE factor = 2.5
Voxel size 3.75 * 3.75 * 4 mm3 3 * 3 * 7 mm3
Field of view (FOV) 24 * 24 cm2 24 * 24 cm2
Number of slices 36 17
Echo time 10.5 ms 17 ms
Repetition time 4 600 ms 4 000 ms
Flip angle 111° 90°
Labeling duration 1 450 ms 1 650 ms
Post-labeling delay 1 525 ms 1 525 ms
Labeling plane (distance from AC-PC line a in head-feet direction) 89 mm 72 mm
Background suppression Yes Yes
Vascular crushers No No
Total scanning time 2:01 min 2:08 min
NEX (no. of excitations) GE; NSA (no. of repetitions) Philips 1 16

a anterior commissure–posterior commissure line