Table 4.

Comparison of the diagnostic performances of the iPhone and LCD monitor

Raters Area under the ROC curve Difference between areas p value 95 % CI
iPhone 5 LCD
1 0.956 0.976 −0.020 0.475 −0.074, 0.035
2 0.955 0.983 −0.027 0.143 −0.064, 0.009
3 0.962 0.983 −0.021 0.150 −0.051, 0.008
4 0.987 0.964 0.024 0.473 −0.041, 0.088
5 0.949 0.957 −0.009 0.726 −0.057, 0.040
6 0.989 0.986 0.003 0.475 −0.006, 0.012
7 0.989 0.954 0.035 0.081 −0.004, 0.075
8 0.981 0.987 −0.006 0.541 −0.026, 0.014
9 0.985 0.993 −0.008 0.326 −0.025, 0.008
10 0.992 0.978 0.014 0.573 −0.036, 0.064
11 0.992 0.993 −0.001 0.746 −0.008, 0.006
12 0.960 0.921 0.039 0.403 −0.053, 0.130
13 0.997 0.988 0.009 0.344 −0.010, 0.027
14 0.981 0.950 0.031 0.479 −0.055, 0.117
15 0.985 0.993 −0.008 0.326 −0.025, 0.008
16 0.992 0.982 0.009 0.668 −0.034, 0.052
Overall, mean (CI) 0.978 (0.965, 0.991) 0.974 (0.960, 0.988) 0.004 0.552 −0.009, 0.017

ROC receiver operating characteristics, LCD liquid crystal display, CI 95 % confidence interval