Skip to main content
. Author manuscript; available in PMC: 2015 Jul 15.
Published in final edited form as: Biopharm Drug Dispos. 2012 Sep 4;33(7):378–402. doi: 10.1002/bdd.1803

Table 7.

Desired and achievable in vitro two-phase parameters to make dissolution test physiological for valproic acid, ondansetron, ibuprofen and felodipine based on in vivo properties and performance

In vivo properties/
performance
In vitro drug
properties
Desired in vitro
parameters
Achievable in vitro parameters



F a a kaa
× 104
Doseb V c Dose
/V
clogD
6.5d
Dae
× 106
PIf ×
104
kpg ×
104
AI/
Vah
F o,
i
M T /
V a j
Vessel
capacity
Depth
below
impellerk
Vessel
AI/Val
V a m kpn ×
104
V o o β Dosep

Actual Decon. Theo. Est. Est. Calc. Est. Est. Based
on ka
Est. Based
on
dose/V
Based on
USP
standard
Req. Est. Req. Req.

Drug name s−1 mg ml mg/ml - cm2/s cm/s s−1 cm−1 - mg/ml ml cm cm−1 ml s−1 ml - mg
Valproic acid (BCS II) 1 11.5 250 100 2.50 1.43 8.87 22.1 11.5 0.52 0.98 2.50 100 1.0 0.47 27 10.4 50 0.02 68
Ondansetron (BCS I or III) 1 5.3 8 100 0.08 1.65 6.63 18.4 5.3 0.29 0.98 0.08 1000 (USP II) 2.5 0.26 302 4.8 338 0.02 24
Ibuprofen (BCS II) 1 7.4 200 100 2.00 2.19 7.50 23.6 7.4 0.31 0.98 2.00 1000 (USP II) 2.5 0.26 302 6.1 97 0.02 604
Felodipine (BCS II) 1 12.5 5 100 0.05 3.41 6.10 20.3 12.5 0.62 0.98 0.05 Not possible unless depth below impeller is decreased to 2 cm and smaller vessel used
a

From reference [36].

b

Arbitrarily chose one of the marketed oral unit doses.

c

Average human fasted intestinal volume.

d

Determined using reference [26].

e

Estimated using Hayduk-Laudie (H-L) method.

f

Estimated using Equation (10) assuming ha = ho = 30μm.

g

Equal to ka.

h

Calculated using Equation (24).

i

Value should ideally be equal or close to Fa in vivo. 0.98 was chosen due to vessel size constraints.

j

Equal to Dose/V.

k

The standard set-up for distance from bottom of vessel to bottom of impeller is 2.5 cm for the 1000 ml USP 2 vessel. 1 cm chosen for 100 ml vessel, which should accommodate a tablet or capsule.

l

From Table 1. Value set by apparatus geometry and minimum or maximum practical volume of media.

m

Volume required to achieve AI/Va.

n

Equal to estimated PI times the actual AI/Va.

o

Volume required to achieve desired Fo,∞ based on cLogD 6.5 and Va.

p

Equal to desirable MT/Va* achievable Va.