Skip to main content
. 2015 May 26;15(6):12377–12388. doi: 10.3390/s150612377

Table 1.

Concentrations of SO42 in pore water samples measured by Raman spectroscopy and liquid chromatography.

Samples Sampling Depth (cm) Concentrations of SO42− (mmol/L) Relative Deviation │CRaman − CLC│/CLC
CRaman CLC
A-12-a 20–60 27.1 28.5 4.91%
B-12-a 60–100 26.1 23.4 11.54%
C-12-a 100–140 25.1 23.6 6.36%
D-12-a 140–180 23.1 22.7 1.76%

CRaman concentrations measured by Raman spectroscopy; CLC concentrations measured by liquid chromatography.