Skip to main content
. 2015 May 26;15(6):12377–12388. doi: 10.3390/s150612377

Table 2.

The fitted linear functions and LODs of conventional Raman and LCOF-Raman experimental setup. (QE65000 is employed as spectrometer, and the length of the LCOF is 100 cm.)

Fitted Linear Function LOD (mmol/L)
Conventional Raman setup R* = 0.012C + 0.534 2.15
LCOF-Raman setup R* = 0.016C + 0.027 1.50