Skip to main content
. 2014 Sep 26;119(9):2132–2147. doi: 10.1002/2013JE004514

Table 1.

Results of REMS-H Reference Model Measurements at Finnish Center of Metrology and Accreditation in a

Ref. RH
Ref. T Ref. RH. uncertainty Mean RH RMS Error
[°C] (% RH) (% RH) (% RH) (% RH)
9.96 7.2 0.2 8.8 1.2
10.06 28.4 0.6 29.8 1.1
10.07 61.5 1.1 55.7 4.1
10 90.6 1.6 85.0 3.3
−70.26 18.4 2 12.3 4.1
−70.3 49 3.5 45.7 3.2
−70.51 94.6 5 92.7 3.7
−59.13 19.6 1.5 19.1 1.1
−59.13 58.6 3 55.5 3.0
−60.14 92.5 3.5 90.5 2.8
−49.95 10.4 1 11.3 1.2
−49.98 30.8 1.5 31.7 1.4
−49.98 60.1 2.5 58.1 2.1
−49.97 92.4 3.5 88.1 3.8
−39.92 9.7 0.6 11.2 1.3
−39.9 29.8 1.5 32.0 2.0
−39.91 60.4 2.5 60.5 1.8
−39.89 90 3 85.4 3.8
−24.77 9.5 0.6 13.0 2.7
−24.85 30.2 1 34.4 3.2
−24.88 60.6 2 62.1 1.8
−24.88 90.3 3 87.4 2.9
−9.82 10.2 0.5 14.3 3.1
−9.79 30 0.9 35.3 3.9
−9.77 60.6 1.7 63.1 2.2
−9.88 91.9 2.5 92.6 2.0
a

The reference temperature, humidity, and uncertainty are provided by the Center and are traceable to national standards. Mean RH is the mean of three Humicap®; readings.