. 2015 Jul 23;42(8):4783–4795. doi: 10.1118/1.4926559

TABLE III.

Absolute dose discrepancies between accumulated and measured doses (values normalized by the local measured doses). Both planning 4D-CT dose accumulations (shown as column “planning 4D-CT”) and on-board 4D-CBCT dose accumulations (shown as column “on-board 4D-CBCT”) were performed and compared to the OSL measurements. The “OSL No.” indicates individual OSL measurements at locations numbered in Fig. 4. Please note that the end-inspiration phase location was fixed for different motion amplitudes. When motion amplitude increased, doses at different OSL measurement locations were affected differently.

Absolute dose discrepancy [absolute (accumulated-measured)/measured*100]
3D static beam Dynamic conformal arc IMRT
OSL No. Planning 4D-CT (%) On-board 4D-CBCT (%) Planning 4D-CT (%) On-board 4D-CBCT (%) Planning 4D-CT (%) On-board 4D-CBCT (%)
3 cm on-board motion 6 6.2 5.5 2.3 1.4 2.2 2.9
9 4.1 3.7 1.4 0.2 0.3 2.5
11 3.3 0.6 3.1 1.2 9.2 5.5
12 5.1 1.6 3.8 0.2 13.8 11.2
15 63.6 2.2 50.6 6.5 43.4 2.4
19 3.7 2.3 0.0 0.6 5.1 4.8
21 3.5 2.0 0.5 0.9 4.5 5.1
Mean 12.8 2.5 8.8 1.6 11.2 4.9
SD 22.4 1.6 18.5 2.2 14.9 3.0
4 cm on-board motion 6 4.0 1.0 3.2 2.2 1.5 1.8
9 7.2 4.5 3.3 1.1 1.7 6.9
11 48.5 1.5 36.3 5.8 68.6 9.6
12 76.2 9.1 66.8 3.8 96.9 1.2
15 105.1 4.8 90.5 4.0 87.9 7.0
19 4.6 0.4 2.9 0.4 2.8 4.7
21 6.8 2.7 3.0 2.0 2.1 3.1
Mean 36.1 3.4 29.4 2.8 37.3 4.9
SD 41.3 3.0 36.4 1.9 44.8 3.1