Skip to main content
. 2014 Nov 15;204(4):515–525. doi: 10.1007/s00430-014-0364-z

Table 3.

Analysis of assay detection rates at 5–25 IU/mL nominal concentration of PEI and WHO replicates

Assay Number of detected samples (IU/mL) Overall detection rate (%)
25 10 5
RealTime 6/6 6/6 4/6 89
CTM v1 6/6 6/6 2/6 78
HPS 6/6 4/6 2/6 67
Versant 6/6 3/6 1/6 56
artus 5/6 2/6 2/6 50
CTM v2 5/6 1/6 1/6 39