Table 6.

Range of uncertainty characteristics

100 IU/mL 1,000 IU/mL
Ref. value LL UL Ref. value LL UL
artus 355 225 836 4,074 2,890 6,900
Versant 102 68 210 1,202 830 2,182
RealTime 74 57 105 724 580 964
CTM v1 186 126 358 1,950 1,480 2,857
CTM v2 162 106 344 1,820 1,373 2,699
HPS 100 66 208 1,000 729 1,592

LL lower limit, UL upper limit; all numbers in IU/mL