Table 2.

Summary of CDE comparisons between Centurion and Infiniti phacoemulsification systems by an individual surgeon and all surgeons combined

Surgeon Infiniti Centurion Difference (std err): Centurion–Infiniti Percent difference P-value*
1 CDE, seconds: mean (std dev) 7.790 (6.669) 5.536 (4.458) −2.254 (0.797) 28.9% 0.005
[range] [1.23, 62.83] [0.23, 27.62]
2 CDE, seconds: mean (std dev) 14.751 (10.225) 9.785 (7.648) −4.966 (0.733) 33.7% <0.001
[range] [2.56, 91.86] [0.08, 61.40]
3 CDE, seconds: mean (std dev) 9.071 (6.611) 5.672 (4.667) −3.399 (0.399) 37.5% <0.001
[range] [0.10, 51.50] [0.07, 48.56]
4 CDE, seconds: mean (std dev) 27.688 (14.510) 15.153 (9.512) −12.535 (1.442) 45.3% <0.001
[range] [5.77, 109.00] [2.16, 84.57]
5 CDE, seconds: mean (std dev) 12.664 (5.394) 7.739 (3.648) −4.926 (0.711) 38.9% <0.001
[range] [4.36, 39.22] [2.58, 23.59]
All CDE, seconds: mean (std dev) 13.330 (10.942) 8.359 (7.200) −4.971 (0.408) 37.3% <0.001
[range] [0.10, 109.00] [0.07, 84.57] −5.088 (0.336)** 38.2% <0.001

Notes:

*

Statistical comparison between patients using Infiniti phacoemulsification system and those using Centurion phacoemulsification system; two-sample t-tests for continuous variables and chi-square tests for categorical variables.

**

Estimated P-value based on linear mixed-effects model, adjusting for age and sex of patients and different surgeons.

Abbreviations: CDE, cumulative dissipated energy; std err, standard error; std dev, standard deviation.