. 2015 Jul 2;42(8):4449–4459. doi: 10.1118/1.4923179

TABLE III.

The WEPL errors between the CTref and CBCT images measured in five angular WEPL profiles for each image.

Absolute error (mean ± SD, mm) Absolute error (mean ± SD, %)a
Phantom CBCTraw CBCTus CBCTap CBCTraw CBCTus CBCTap
Head (L) 12.4 ± 9.4 3.6 ± 3.7 1.5 ± 2.0 14.3 ± 5.6 5.1 ± 4.7 1.9 ± 2.4
Head (H) 10.4 ± 8.0 4.6 ± 5.3 1.3 ± 1.8 11.9 ± 4.8 5.6 ± 5.1 1.8 ± 2.1
Thorax (L) 6.6 ± 3.5 6.6 ± 6.4 3.3 ± 2.6 8.3 ± 5.7 7.6 ± 7.4 3.6 ± 3.0
Thorax (H) 6.1 ± 3.4 7.8 ± 6.7 3.0 ± 2.4 7.5 ± 4.9 9.2 ± 8.6 3.4 ± 2.8
Abdomen (L) 5.0 ± 4.9 22.4 ± 9.4 2.2 ± 1.4 3.5 ± 2.4 19.3 ± 6.4 1.9 ± 1.0
Abdomen (H) 5.0 ± 4.9 14.6 ± 7.0 2.0 ± 1.7 3.5 ± 2.4 13.1 ± 5.4 1.7 ± 1.3
Pelvis (L) 6.8 ± 4.4 12.8 ± 6.7 2.8 ± 1.9 6.2 ± 4.4 12.1 ± 8.2 2.2 ± 1.4
Pelvis (H) 5.8 ± 4.3 16.4 ± 7.6 2.8 ± 1.9 5.2 ± 4.4 15.2 ± 9 2.2 ± 1.4
Overall 7.3 ± 5.3 11.1 ± 6.6 2.4 ± 2.0 7.6 ± 4.3 10.9 ± 6.9 2.3 ± 1.9
a

Percentage error relative to the WEPL of CTref; H = high mAs; L = low mAs; SD = one standard deviation.