. 2015 Jul 2;42(8):4449–4459. doi: 10.1118/1.4923179

TABLE IV.

The 3D gamma analysis result for proton treatment plans on the phantoms. The numbers represent the percentages of voxels passing the gamma evaluation.

Gamma criteria: 3 mm/3% Gamma criteria: 2 mm/2%
Plan CBCTraw CBCTus CBCTap CBCTraw CBCTus CBCTap
Head 85.6 87.7 99.5 78.6 78.1 98.1
Thorax (lung) 84.4 69.0 98.6 79.9 58.6 93.0
Thorax (mediastinum) 96.4 97.5 99.7 91.6 94.4 98.8
Pelvis 94.8 92.4 98.4 85.5 81.7 94.3
Overall 90.3 ± 5.3 86.7 ± 10.8 99.0 ± 0.6 83.9 ± 5.1 78.2 ± 12.8 96.0 ± 2.5